Nature of Requirements:
ADVANCE CONTRACT AWARD NOTICE (ACAN)
AIRPOINTER
K3D57-160675/A
1. Advance Contract Award Notice (ACAN)
An ACAN is a public notice indicating to the supplier community that a department or agency intends to award a contract for goods, services or co
nstruction to a pre-identified supplier, thereby allowing other suppliers to signal their interest in bidding, by submitting a st
atement of capabilities. If no supplier submits a statement of capabilities that meets the requirements set out in the ACAN, on or before the closing date stated in the ACAN, the contracting officer may then
proceed with the award to the pre-identified supplier.
2. Definition of the requirement
2.1 Background:
Environment Canada (EC) - Science and Technology Branch requires the supply and delivery of two (2) complete and fully assembled PM Air Pointer instruments and accessories - one (1) 03NOx Airpointer and one (1) H2S/SO2 Airpointer. Each Airpointer will be used to support resear
ch to measure ozone, hydrogen sulfide, sulfur dioxide, nitrogen oxides and PM2.5 in a single, self-contained, temperature-controlled unit that can be l
eft unattended in the field under harsh conditions for many months. These systems are to be used to quantify the contribution of increasing sh
ip traffic, as well as local sources, on air pollution in Arctic communities, as well as for other projects involving air quality measurements across Canada.
2.2 General Information
2.2.1 The required H2S/SO2 system consists of one (1) AirPointer PM Base, equipped with one (1) for each of the following: H2S / SO2 Analyzer, *BTEX GC, SO2 Permeation Device, H2O Permeation Device, SO2 Internal Span, SO2/H2S Int. Span, WS/WD/T/RH/P/R, and Wireless 3G.
2.2.2 The required03/NOx system consists of one (1) AirPointer PM Base
, equipped with one (1) for each of the following: O3 Analyzer, NOx Analyzer, SO2 Analyzer, SHARP Model 5030, O3 Internal Span, NO2 In
ternal Span, NO2 Permeation Device, SO2 Internal Span, SO2 Permeation Device, WS/WD/T/RH/P/R, Stand
3. Mandatory System Requirements
3.1 Minimum Essential Requirements:
Any interested supplier must demonstrate by way of a statement of capabilities that is products meets the following specifications:
1. Variables measured, range, precision required:
a. O3: 0-1000 ppb, 1 ppb;
b. NOx: 0-1000 ppb, 2 ppb;
c. H2S/SO2: 0-1000 ppb, 1ppb;
e. PM2.5: 0-100 ug/m3, 1 ug/m3, EPA equivalent 6.
Meteorology: T, RH, WS, WD, P.
2. Data time resolution: 1 minute.
3. Fully integrated data logger and computer which collects all data plus sensor pressures, temperatures, flow rates, with Ethernet, USB, VGA.
4. Automated zero for all gases (i.e internal zero air generator).
5. Automated span calibration for H2S/SO2/O3/NO2.
6. Fully weatherproofed housing. Ambient Operating T range: -40C to +40C.
7. Actively temperature controlled box (thermostat, AC & heater).
8. T-controlled instrument box dimensions