Natural Resources Canada (NRCan) has a requirement for the supply of a High Pressure Ion Chromatograph System and Components. The requirement must include all of the following specifications and components below.Specifications and Components: A complete Ion Chromatography system with metal free pathway that supports standard bore and microbore applications.A midual pump for analytical gradients, analytical isocratic runs. Must also have integrated sealed in line degassing and mixing of up to four (4) eluents.Autosampler with a minimum of three separate trays of 1.5 ml samples and full temperature control.Must have two separate zone column compartment with two (2) injection valves. A third (3rd) two (2)-way six (6) port valve and mounting hardware must be included for more flexibility.Four (4) channel variable wavelength detector with eleven (11) microliter (uL) peek analytical flowcellElectrolytic eluent generation.A conductivity detector.A electrochemical detector and cell.A suppressor device with the ability to suppress all of the following eluents: carbonate, hydroxide, or methanesulfonic acid eluents as required for EPA, ASTM, ISO, or other standardized methods.A complete software package with PC, monitor included as well as all licensing required.A complete package with columns and guards for Carbohydrates as well as standard anions and cation analysis. System must be capable of running 2 um (micron) particle size columns.