Trade Agreement: NAFTA/AIT/Canada FTAs with Peru/Colombia/Panama
Tendering Procedures: All interested suppliers may submit a bid
Competitive Procurement Strategy:
Comprehensive Land Claim Agreement: No
Nature of Requirements:
PORTABLE DATA ACQUISITION SYSTEM
1. Advance Contract Award Notice (ACAN)
An ACAN is a public notice indicating to the supplier community that a department or agency intends to award a contract for goods, services or construction to a pre-identified supplier, thereby allowing other suppliers to signal their interest in bidding, by submitting a statement of capabilities. If no supplier submits a statement of capabilities that meets the requirements set out in the ACAN, on or before the closing date stated in the ACAN, the contracting officer may then proceed with the award to the pre-identified supplier.
2. Definition of the requirement
Public Works and Government Services Canada (PWGSC), on behalf of its client, Defence Research and Development Canada (DRDC), is proposing to negotiate the supply of a Portable Data Acquisition System.
DRDC own a Dewetron DEWE2-A7 Portable Data Acquisition System to update instrumentation in its laboratories. In order to continue to update this instrumentation, DRDC wishes to acquire a second system identical or equivalent to DEWE2-A7. The new system must be compatible with the Dewetron DEWE2-A7 and be commercially available (COTS).
The goal of having two identical systems will serve to:
- Have a bank of modules for both systems.
- Connect these two systems in order to have a system with remote inputs (which reduces the length of cable between the sensor and the acquisition system).
The two acquisition systems will be used jointly in different laboratories for data acquisition during tests of :
- Rocket engine with static pressure, dynamic pressure, forces, temperatures.
- Measurement of dynamic pressures in rifle barrels.
- Measurement of forces and pressures in a tri-sonic wind tunnel.
Here are some types of sensors used:
- ICP Piezoelectric Pressure Sensor
- Pressure sensor with strain gauges
- ICP piezoelectric force sensor
- Thermocouples different types K, T, J
- Rotary Encoder
- Reading of different status sensors (TTL or 24vdc)
3. Criteria for assessment of the Statement of Capabilities (Minimum Essential Requirements)
Any interested supplier must demonstrate by way of a statement of capabilities that its product meets the requirements of the requirement and that it includes the following components and/or specifications:
Quantity required: 1 of each
Portable all-in-one data acquisition system, rugged industrial design
- 7slots for user exchangeable modules
- 1x PCIe 1X interface
- 17"? full HD display, 1920 x 1080 pixels *(or an external port to connect a 17’’ full HD display)
- Intel® CoreTM i7 processor, 4 GB RAM
- 480 GB flash disk no moving parts, max. data throughput 100 MB/s
- 7x USB, 1x LAN Ethernet, , 2x Firewire 1394a
- 1x external DVI socket, AUDIO interface
- Option to link with other DEWE2 system
- Power supply 115 / 230 VAC
- Microsoft® Windows® 7 operating system
- With Acquisition software
- Battery power supply with isolated DC
- 2 hours operation included
6 channel module with isolated inputs for constant current (ICP®) sensors and voltage
- 16-bit resolution, simultaneous sampling, 2 MS/s per channel
- Freely programmable 2nd, 4th, 6th order low-pass filters 0.16 Hz high-pass filter
- Measure modes: IEPE, voltage and current (via external shunt)
- 14 programmable voltage ranges from ±5 mV to ±100 V
- 4 / 8 mA IEPE constant current supply
- Isolation 1500 V
- Self-check functionality
- Connectors: isolated BNC sockets
8 channel universal analog input module
- 24-bit resolution, simultaneous sampling, 204.8 kS/s per channel, anti-aliasing filters
- Bandwidth DC to 77 kHz
- Measure modes: bridge, voltage, IEPE, resistance, temperature and current
- Programmable excitation voltage from 0 - 13 VDC, independent for each channel
- Programmable excitation current from 200 ìA - 25 mA, independent for each channel
- DC or AC coupling, 4 high pass filters 0.16, 0.5, 3.4 and 10 Hz
- Bridge input: full bridge sensors 80 Ohm to 10 kOhm (@ excitation