M1 The automated sample dilution system must be capable of performing dilutions up to 2000x
M2 The automated sample dilution system must be capable of performing serial dilutions, standard additions, spikes, and sample splitting
M3 The automated sample dilution system must have at least 4 rack positions to accommodate up to 240 test tubes for overnight operation (using 4x 60-
tube racks)
M4 The automated sample dilution system must include a high purity syringe for use with samples containing Nitric acid or Hydrochloric acid
M5 Measured carryover of the automated sample dilution system between samples must be 0.9999
M9 The automated sample dilution system must be capable of handling acidified samples
M10 The automated sample dilution system must be compatible with syringes between 1.0mL – 50mL
M11 The automated sample dilution system must include software to program sample dilution sequences for each system purchased.
M12 The automated sample dilution system must have the capability to function as an autosampler that is compatible with Thermo ICAP/OES instruments
M13 The automated sample dilution system must be able to sit atop a countertop with dimensions of 30”W, 36”D, 60”H