1. Advance Contract Award Notice (ACAN)
An ACAN is a public notice indicating to the supplier community that a department or agency intends to award a contract for goods, services or construction to a pre-identified supplier, thereby allowing other suppliers to signal their interest in bidding, by submitting a statement of capabilities. If no supplier submits a statement of capabilities that meets the requirements set out in the ACAN, on or before the closing date stated in the ACAN, the contracting officer may then proceed with the award to the pre-identified supplier.
2. Definition of the requirement
The Tobacco Control Directorate’s Science Division of Health Canada has a requirement for the supply, installation, training and manuals for a hybrid Liquid Chromatography (LC) Mass Spectrometer (MS) Instrument. The Science Division requires LC-MS instrument to establish analytical capabilities for the testing of vaping and other novel tobacco products.
3. Criteria for assessment of the Statement of Capabilities (Minimum Essential Requirements)
Any interested supplier must demonstrate by way of a statement of capabilities that its product/equipment/system (as appropriate) meets the following requirements:
A. LIQUID CHROMATOGRAPH (LC)
- LC must be capable of ultra high performance liquid chromatography and capable of generating pressure up to 15000 psi for fast chromatography analysis;
- LC must be capable of performing high-throughput sample analysis;
- LC must be equipped with a binary pump capable of high pressure gradient mixing;
- LC column compartment must be capable of heating up to 120°C;
- LC sample compartment must be capable of sample refrigeration cooling at 4°C;
- Autosampler must be able to accommodate trays for standard 2 mL autosampler vials; and
- LC system capable of high precision at low volume (1µL) injection.
B. MASS SPECTROMETER (MS)
- Minimum instrument’s mass range: 5-3000 m/z
- MS must be capable of mass resolution of 70,000 at m/z 200 scan rate
- MS must have mass accuracy