Trade Agreement: NAFTA/Canada FTAs with Peru/Colombia/Panama
Tendering Procedures: All interested suppliers may submit a bid
Attachment: None
Competitive Procurement Strategy: Lowest/Lower Bid
Comprehensive Land Claim Agreement: No
Nature of Requirements:
LOAD FRAME EXTENSOMETER CALIBRATION
W7707-145698/A
Gillis, Floyd
Telephone No. - (902) 496-5566
Fax No. - (902) 496-5016
1. Background:
Defense Research/Development Canada- Atlantic (DRDC/A) has a
requirement for annual calibration (for 3 years) of load frames
and extensometers corresponding ASTM standards (Appendix A) by a
qualified technician accredited in according to American
Association for Laboratory Accreditation A2LA or ISO/IEC 17025
for the purpose of primarily running ASTM E8M tests.
2. Scope:
Annual calibration for three ( 3) years of servo-hydraulic and
screw driven load frames and accompanying extensometers (listed
below) with corresponding ASTM standards.
3. Requirement:
The following load frames and extensometers equipment shall be
calibrated annually for three (3) years:
1 MN Load frame, Instron 1335 SN H0012
· LVDT model 2411-186, SN 3306-64B, range ±150 mm
· Load cell model 2518-120, SN UK024, range ±1 MN, 3.5"x12 TPI
threads
2. 250 kN Load frame, MTS Landmark 370.25 SN 10313986
· LVDT model 370.25, SN 848604, range ±150 mm
· Load cell model 661.21H-01, SN 10306870D, range ±250 kN, m36x2
mm threads
3. 25 kN Load frame, MTS Landmark 370.02 SN 10313996
· LVDT model 370.02, SN 849934, range ±100 mm
· Load cell model 661.19H-04, SN 108517A, range ±25 kN, m12x1.25
mm threads
· Load cell model 661.19H-02, SN 108517A, range ±10 kN, m12x1.25
mm threads
4. 5 kN Load frame, United SSTM-5KN SN 1209539
· LVDT SN 1209539, range ±1 m
· Load cell SN T1I106793, range ±5 kN, 5/8" socket ¼" hitch pins
· Built in extensometer model EL-20, SN 5161, range 1 m
5. Static Strain Extensometer
· Model 634.12F-24, SN 10304709A, range ±0.5 mm
· Model 634.12F-24, SN 10304709C, range ±0.5 mm
· Model 634.12F-24, SN 10312611A, range ±4 mm
· Model 632.13F-24, SN 10311008C, range ±4 mm
· Model 632.13F-20, SN 10310281A, range ±0.15 mm
· Model 632.13F-20, SN 10312855, range ±0.15 mm
6. Clip Gauge Extensometers
· Model 632.03F-30, SN 10312800, range ±4 mm
· Model 632.03F-20, SN 10305679D, range ±12 mm
All listed equipment will be calibrated to:
All LVDTs shall be calibrated to ±1.0 % of applied length from
10% - 100% of range, and ± 0.1% of range below 10%
All Load Cells shall be calibrated in compliance with ASTM E4-10
(or newer version)
All Extensometers shall be calibrated in compliance with ASTM
E83-10 (or newer version)
All Clip gauges or length devices shall be calibrated in
compliance with ASTM E399-10 (or newer version)
All reference standards shall be calibrated in compliance with
ASTM E74
All calibration shall be carried out on site at Dockyard
Laboratory (Atlantic) located in Bldg D-20, HMC Dockyard, CFB
Halifax, Halifax, NS.
Work must be carried out by a qualified technician accredited to
American Association for Laboratory Accreditation A2LA or
ISO/IEC 17025.
The Request for Proposal (RFP) document forms the solicitation
requirement.
This office provides service to the public in English.
Delivery Date: Above-mentioned
The Crown retains the right to negotiate with suppliers on any
procurement.
Documents may be submitted in either official language of Canada.