Trade Agreement: WTO-AGP/NAFTA/AIT/Canada-Peru
FTA/Canada-Colombia FTA
Tendering Procedures: Generally only one firm has been invited
to bid
Attachment: None
Non-Competitive Procurement Strategy: Exclusive Rights
Comprehensive Land Claim Agreement: No
Vendor Name and Address:
SCL Instrumentation Inc.
205-48 Laverock Avenue
Richmond Hill Ontario
Canada
L4C4J5
Nature of Requirements:
Air Quality Monitoring System Airpointer
KM061-121140/A
Juan, Peggy
Telephone No. - (905) 615-2467 ( )
Fax No. - (905) 615-2060
1. ACAN Explanation
An Advance Contract Award Notice (ACAN) allows departments and
agencies to post a notice, for no less than fifteen (15)
calendar days, indicating to the supplier community that it
intends to award a good, service or construction contract to a
pre-identified contractor. If no other supplier submits, on or
before the closing date, a Statement of Capabilities that meets
the requirements set out in the ACAN, the contracting authority
may then proceed with the award.
2. Definition of Requirement
Environment Canada, Air Quality Research Division in Toronto,
Ontario, requires the supply and delivery of quantity two (2)
complete and fully-assembled Airpointer Air Monitoring System
base units equipped with SO2 permeation tube, H2S permeation
tube and holding stands. Environment Canada currently owns and
operates three Airpointers and has developed an in-house data
analysis software specific to this instrument and plans to
continue utilizing the same data analysis software without
redevelopment. Delivery of the two units is required on or
before March 31, 2012.
3. Criteria for Assessment of the Statement of Capabilities
3.1 Minimum Essential Requirements:
Any interested supplier must demonstrate by way of a statement
of capabilities that its product meets the following
requirements:
An air quality monitoring system meeting the following
specifications:
1. Variables measured, range, precision required:
a. O3: 0-1000 ppb, 1 ppb;
b. NOx: 0-1000 ppb, 2 ppb;
c. H2S/SO2: 0-1000 ppb, 1 ppb;
d. CO: 0-100 ppm, 0.1 ppm;
e. PM2.5: 0-100 ug/m3, 1 ug/m3, EPA equivalent 6. Meteorology:
T, RH, WS, WD, P.
2. Data time resolution: 1 minute.
3. Fully integrated data logger and computer which collect all
data plus sensor pressures, temperatures, flow rates Ethernet,
USB, VGA.
4. Automated zero for all gases (i.e. internal zero air
generator).
5. Automated span calibration for H2S/SO2.
6. Fully weatherproofed housing. Ambient Operating T range: -40C
to + 40C.
7. Actively temperature controlled box (thermostat, AC &
heater).
8. T-controlled instrument box dimensions